Результаты исследований: Научные публикации в периодических изданиях › статья › Рецензирование
A universal capacitive spectrometer, assembled based on a bridge scheme with successive phase-sensitive detection of an hf signal (10 MHz) and a feedback system, which provides a constant capacitance regime, is described. The spectrometer permits measuring the parameters of deep centers whose relative concentration varies from 10** minus **5 up to 1 in the range of relaxation times of the signal 10** minus **4-1 sec, determined by synchronous detection of the signal, for filling pulse durations of 10 nsec to 1 msec. The spectrometer is simple to calibrate. The minimum time for clearing the feedback system equals 10 mu sec.
| Язык оригинала | английский |
|---|---|
| Страницы (с-по) | 696-701 |
| Число страниц | 6 |
| Журнал | Instruments and experimental techniques New York |
| Том | 30 |
| Номер выпуска | 3 pt 2 |
| Состояние | Опубликовано - мая 1987 |
ID: 75076859