DOI

The microelectronic design and some results of experimental tests of three types of coordinate-sensitive Micro-Channel-Plane (MCP) based detectors with good timing properties are presented. The electrodynamic structures that could be used for the MCP fast signal readout (strips, delay lines and discrete pads) are always a compromise between the amplitude of the induced signal, propagation time delay, cross-talk and distortions of the signal at the output. In-lab tests of these structures and experimental tests of the detectors done with the beams of MIPs and with the alpha-source, have shown a 130 μm spatial resolution and a 110 ps timing resolution obtained simultaneously for one detector.

Язык оригиналаанглийский
Страницы (с-по)402-407
Число страниц6
ЖурналNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Том367
Номер выпуска1-3
DOI
СостояниеОпубликовано - 11 дек 1995

    Предметные области Scopus

  • Ядерная физика и физика высоких энергий
  • Контрольно-измерительные инструменты

ID: 51039178