A Change in the Parameters of Si(Li) Detectors under Exposure to α Particles. / Bazlov, N. V.; Bakhlanov, S. V.; Derbin, A. V.; Drachnev, I. S.; Izegov, G. A.; Kotina, I. M.; Muratova, V. N.; Niyazova, N. V.; Semenov, D. A.; Trushin, M. V.; Unzhakov, E. V.; Chmel, E. A.
в: Instruments and Experimental Techniques, Том 63, № 1, 01.01.2020, стр. 25-29.Результаты исследований: Научные публикации в периодических изданиях › статья › Рецензирование
}
TY - JOUR
T1 - A Change in the Parameters of Si(Li) Detectors under Exposure to α Particles
AU - Bazlov, N. V.
AU - Bakhlanov, S. V.
AU - Derbin, A. V.
AU - Drachnev, I. S.
AU - Izegov, G. A.
AU - Kotina, I. M.
AU - Muratova, V. N.
AU - Niyazova, N. V.
AU - Semenov, D. A.
AU - Trushin, M. V.
AU - Unzhakov, E. V.
AU - Chmel, E. A.
N1 - Funding Information: This work was supported by the Russian Science Foundation (project no. 17-12-01009) and the Russian Foundation for Basic Research (project nos. 16-29-13014, 17-02-00305, and 19-02-0009). Publisher Copyright: © 2020, Pleiades Publishing, Ltd. Copyright: Copyright 2020 Elsevier B.V., All rights reserved.
PY - 2020/1/1
Y1 - 2020/1/1
N2 - Abstract—: A spectrometer based on a Si(Li) detector has been tested by exposure to α particles with the aim of determining changes in the detector characteristics versus the fluence. As a result of 60-day measurements with a total fluence of 6.2 × 109 α particles, it has been established that the dependence of the deterioration in the energy resolution of α-particle peaks on the fluence is described by a linear function with a slope of Δσ/ΔФ = (8.4 ± 0.4) × 10–10 keV/α. The measured amplitude of the α-particle momentum decreases linearly with increasing fluence with a slope of (–4.8 ± 0.6) × 10–9 keV/α. These effects do not prevent reliable discrimination between the signals from α particles and fission fragments until the fluence reaches a value of 1010 α particles when the Si(Li) detector is a part of a neutron source for detecting fission fragments.
AB - Abstract—: A spectrometer based on a Si(Li) detector has been tested by exposure to α particles with the aim of determining changes in the detector characteristics versus the fluence. As a result of 60-day measurements with a total fluence of 6.2 × 109 α particles, it has been established that the dependence of the deterioration in the energy resolution of α-particle peaks on the fluence is described by a linear function with a slope of Δσ/ΔФ = (8.4 ± 0.4) × 10–10 keV/α. The measured amplitude of the α-particle momentum decreases linearly with increasing fluence with a slope of (–4.8 ± 0.6) × 10–9 keV/α. These effects do not prevent reliable discrimination between the signals from α particles and fission fragments until the fluence reaches a value of 1010 α particles when the Si(Li) detector is a part of a neutron source for detecting fission fragments.
UR - http://www.scopus.com/inward/record.url?scp=85081585596&partnerID=8YFLogxK
U2 - 10.1134/S0020441220010029
DO - 10.1134/S0020441220010029
M3 - Article
AN - SCOPUS:85081585596
VL - 63
SP - 25
EP - 29
JO - Instruments and Experimental Techniques
JF - Instruments and Experimental Techniques
SN - 0020-4412
IS - 1
ER -
ID: 75075943