DOI

XPS measurements of coarse-grained and nanostructured nitinol (Ni 50.2Ti49.8) before and after chemical treatment in hydrofluoric acid (40% HF, 1 min) are presented. The nanostructured state, providing the excellent mechanical properties of nitinol, is achieved by severe plastic deformation. The near-surface layers of nitinol were studied by XPS depth profiling. According to the obtained results, a chemical treatment in hydrofluoric acid reduces the thickness of the protective TiO2 oxide layer and induces a nickel release from the nitinol surface and an arsenic contamination, and can therefore not be recommended as conditioning to increase the roughness of NiTi-implants. A detailed evaluation of the resulting toxicological risks is given. © 2012 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2012.

Original languageEnglish
Pages (from-to)1812-1816
Number of pages5
JournalJournal of Biomedical Materials Research - Part B Applied Biomaterials
Volume100 B
Issue number7
DOIs
StatePublished - 1 Oct 2012
Externally publishedYes

    Research areas

  • arsenic contamination, nitinol, TiO, XPS

    Scopus subject areas

  • Biomaterials
  • Biomedical Engineering

ID: 35166922