Research output: Contribution to journal › Article › peer-review
A measurement method of a detector response function for monochromatic electrons based on the Compton scattering. / Bakhlanov, S. V.; Bazlov, N. V.; Derbin, A. V.; Drachnev, I. S.; Kayunov, A. S.; Muratova, V. N.; Semenov, D. A.; Unzhakov, E. V.
In: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 821, 11.06.2016, p. 13-16.Research output: Contribution to journal › Article › peer-review
}
TY - JOUR
T1 - A measurement method of a detector response function for monochromatic electrons based on the Compton scattering
AU - Bakhlanov, S. V.
AU - Bazlov, N. V.
AU - Derbin, A. V.
AU - Drachnev, I. S.
AU - Kayunov, A. S.
AU - Muratova, V. N.
AU - Semenov, D. A.
AU - Unzhakov, E. V.
N1 - Funding Information: This work was supported by the Russian Foundation for Basic Research (Projects 15-02-02117-A , 13-02-01199-A and 13-02-12140-ofi-m ). Publisher Copyright: © 2016 Elsevier B.V. Copyright: Copyright 2017 Elsevier B.V., All rights reserved.
PY - 2016/6/11
Y1 - 2016/6/11
N2 - In this paper we present a method of scintillation detector energy calibration using the gamma-rays. The technique is based on the Compton scattering of gamma-rays in a scintillation detector and subsequent photoelectric absorption of the scattered photon in the Ge-detector. The novelty of this method is that the source of gamma rays, the germanium and scintillation detectors are immediately arranged adjacent to each other. The method presents an effective solution for the detectors consisting of a low atomic number materials, when the ratio between Compton effect and photoelectric absorption is large and the mean path of gamma-rays is comparable to the size of the detector. The technique can be used for the precision measurements of the scintillator light yield dependence on the electron energy.
AB - In this paper we present a method of scintillation detector energy calibration using the gamma-rays. The technique is based on the Compton scattering of gamma-rays in a scintillation detector and subsequent photoelectric absorption of the scattered photon in the Ge-detector. The novelty of this method is that the source of gamma rays, the germanium and scintillation detectors are immediately arranged adjacent to each other. The method presents an effective solution for the detectors consisting of a low atomic number materials, when the ratio between Compton effect and photoelectric absorption is large and the mean path of gamma-rays is comparable to the size of the detector. The technique can be used for the precision measurements of the scintillator light yield dependence on the electron energy.
KW - Energy scale calibration
KW - Scintillator and semiconductor detectors
UR - http://www.scopus.com/inward/record.url?scp=84962601248&partnerID=8YFLogxK
U2 - 10.1016/j.nima.2016.03.054
DO - 10.1016/j.nima.2016.03.054
M3 - Article
AN - SCOPUS:84962601248
VL - 821
SP - 13
EP - 16
JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
SN - 0168-9002
ER -