Результаты исследований: Научные публикации в периодических изданиях › статья › Рецензирование
An ion-catcher device consisting of a buffer-gas stopping cell and a radio-frequency quadrupole (RFQ) has been built for the SHIPTRAP facility at GSI. Results of characterisation measurements with the buffer-gas cell and the extraction RFQ performed at GSI in Darmstadt and at the MLL (Maier-Leibnitz- Laboratory) in Garching are presented. The set-up was tested off-line using laser-produced ions and on-line using stable beams and fusion-evaporation products. During the on-line measurements the ions with total energies of around 200 keV/u were thermalised in helium buffer gas at 40-60 mbar. In the following they were guided by a combination of electric RF- and DC-fields until they were transported by the gas flow through the extraction nozzle. After being extracted by a supersonic gas jet the ions were separated from the buffer gas and guided by the extraction RFQ towards subsequent detection systems. Depending on the electric-field strength average extraction times of around 10 ms and an overall efficiency (including stopping and extraction) between 4% and 8% have been achieved.
Язык оригинала | английский |
---|---|
Страницы (с-по) | 489-500 |
Число страниц | 12 |
Журнал | Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms |
Том | 244 |
Номер выпуска | 2 |
DOI | |
Состояние | Опубликовано - 1 мар 2006 |
ID: 37416123