DOI

A standardless electrochemical method for metal ion determination in aqueous solution is proposed. This method is based on the combination of stripping voltammetry and controlled potential coulometry laws. Its application requires neither calibration using standard solutions nor standard additions to the studying solutions. The efficiency of the new method was examined using the detection of mercury, cadmium, copper and lead ions in aqueous media. The interference of metal ions was studied, and the determination of each metal ion was carried out in the presence of the others. The method was tested on samples of river water analyzed independently by atomic absorption spectrometry. A good agreement of the results was demonstrated.

Язык оригиналаанглийский
Страницы (с-по)2222-2226
Число страниц5
ЖурналElectroanalysis
Том19
Номер выпуска21
DOI
СостояниеОпубликовано - 1 ноя 2007

    Предметные области Scopus

  • Аналитическая химия
  • Электрохимия

ID: 36705174