Overloading control of gas chromatographic systems. / Zenkevich, I.G.; Pavlovskii, A.A.
в: Journal of Separation Science, Том 38, № 16, 2015, стр. 2848-2856.Результаты исследований: Научные публикации в периодических изданиях › статья › Рецензирование
}
TY - JOUR
T1 - Overloading control of gas chromatographic systems
AU - Zenkevich, I.G.
AU - Pavlovskii, A.A.
PY - 2015
Y1 - 2015
N2 - Increasing the interlaboratory reproducibility of gas chromatographic retention indices requires avoiding measurements distorted by overloading effects. Several criteria of evaluating the limits of the mass overloading of gas chromatographic systems are compared and reconsidered. The criteria mostly appropriate for practical purposes are based on (i) the dependences of factors of peak broadening (ratio of peak height and its width) vs. amount of analyte injected into the chromatographic column and (ii) the dependence of parameters characterizing the peak distortion (asymmetry factor) vs. the amount of analyte. Both these criteria provide mutually comparable evaluations of the overloading limits for analytes of different polarity. At the same time, the dependence of retention indices vs. amounts of analyte injected in the chromatographic column cannot be recommended for overloading control, because the parameters of the corresponding linear regressions indicate temperature dependence. The interpretation of cer
AB - Increasing the interlaboratory reproducibility of gas chromatographic retention indices requires avoiding measurements distorted by overloading effects. Several criteria of evaluating the limits of the mass overloading of gas chromatographic systems are compared and reconsidered. The criteria mostly appropriate for practical purposes are based on (i) the dependences of factors of peak broadening (ratio of peak height and its width) vs. amount of analyte injected into the chromatographic column and (ii) the dependence of parameters characterizing the peak distortion (asymmetry factor) vs. the amount of analyte. Both these criteria provide mutually comparable evaluations of the overloading limits for analytes of different polarity. At the same time, the dependence of retention indices vs. amounts of analyte injected in the chromatographic column cannot be recommended for overloading control, because the parameters of the corresponding linear regressions indicate temperature dependence. The interpretation of cer
KW - Gas chromatography
KW - Overloading control
KW - Overloading criteria
U2 - 10.1002/jssc.201401471
DO - 10.1002/jssc.201401471
M3 - Article
VL - 38
SP - 2848
EP - 2856
JO - Journal of Separation Science
JF - Journal of Separation Science
SN - 1615-9306
IS - 16
ER -
ID: 3969021