Multivariate calibration transfer between ATR-FTIR and NIR spectrometers: Electronic cigarette refill fluids case study. / Юргенсон, Ника; Монахова, Юлия Борисовна; Кирсанов, Дмитрий Олегович.
в: Microchemical Journal, Том 212, 113375, 05.2025.Результаты исследований: Научные публикации в периодических изданиях › статья › Рецензирование
}
TY - JOUR
T1 - Multivariate calibration transfer between ATR-FTIR and NIR spectrometers: Electronic cigarette refill fluids case study
AU - Юргенсон, Ника
AU - Монахова, Юлия Борисовна
AU - Кирсанов, Дмитрий Олегович
PY - 2025/5
Y1 - 2025/5
N2 - This paper explores an opportunity of transferring multivariate calibration models between the spectroscopic instruments operating on different principles: near infrared spectrometry (NIR) and attenuated total reflection infrared spectrometry (ATR-IR). As an example of such a cross-platform transfer we study the spectroscopic data obtained in the quantification of nicotine, glycerol and propylene glycol in e-cigarette refilling liquids. It is demonstrated that the application of the direct standardization (DS) protocol allowed using the PLS regression models calculated for NIR spectrometer data together with the data obtained with ATR-IR spectrometer and vice versa. The root mean squared error of prediction (RMSEP) values for the transferred data were between 1.0 and 1.5 times higher than that for the original calibrations. The resulted transferred model's accuracy depends a lot on a particular choice of a transfer set and in certain cases the transferred models show comparable accuracy as native models constructed and tested with original instrumental response. This opens up a perspective of developing “universal” cross-platform multivariate calibrations that would be suitable for a particular analytical task regardless the physical operation principles of a measuring device they applied to.
AB - This paper explores an opportunity of transferring multivariate calibration models between the spectroscopic instruments operating on different principles: near infrared spectrometry (NIR) and attenuated total reflection infrared spectrometry (ATR-IR). As an example of such a cross-platform transfer we study the spectroscopic data obtained in the quantification of nicotine, glycerol and propylene glycol in e-cigarette refilling liquids. It is demonstrated that the application of the direct standardization (DS) protocol allowed using the PLS regression models calculated for NIR spectrometer data together with the data obtained with ATR-IR spectrometer and vice versa. The root mean squared error of prediction (RMSEP) values for the transferred data were between 1.0 and 1.5 times higher than that for the original calibrations. The resulted transferred model's accuracy depends a lot on a particular choice of a transfer set and in certain cases the transferred models show comparable accuracy as native models constructed and tested with original instrumental response. This opens up a perspective of developing “universal” cross-platform multivariate calibrations that would be suitable for a particular analytical task regardless the physical operation principles of a measuring device they applied to.
KW - Calibration transfer
KW - Chemometrics
KW - Direct standardization
KW - Electronic cigarettes
KW - Infrared spectroscopy
KW - Near-infrared spectroscopy
KW - Partial least squares regression
UR - https://www.mendeley.com/catalogue/f7666476-790e-300e-9a29-ed19e809991d/
U2 - 10.1016/j.microc.2025.113375
DO - 10.1016/j.microc.2025.113375
M3 - Article
VL - 212
JO - Microchemical Journal
JF - Microchemical Journal
SN - 0026-265X
M1 - 113375
ER -