Standard

Model of the deviation of the medical needle during the movement in human tissue. / Morozov, Victor A.; Nikitin, Sergey A.; Harlamov, Vyacheslav V.; Druzhinin, Vasilii G.

в: Russian Journal of Biomechanics, Том 22, № 4, 01.01.2018, стр. 401-413.

Результаты исследований: Научные публикации в периодических изданияхстатьяРецензирование

Harvard

Morozov, VA, Nikitin, SA, Harlamov, VV & Druzhinin, VG 2018, 'Model of the deviation of the medical needle during the movement in human tissue', Russian Journal of Biomechanics, Том. 22, № 4, стр. 401-413. https://doi.org/10.15593/RJBiomech/2018.4.03

APA

Vancouver

Author

Morozov, Victor A. ; Nikitin, Sergey A. ; Harlamov, Vyacheslav V. ; Druzhinin, Vasilii G. / Model of the deviation of the medical needle during the movement in human tissue. в: Russian Journal of Biomechanics. 2018 ; Том 22, № 4. стр. 401-413.

BibTeX

@article{7bd8c71ec43a4bc6bb9c09888764e48b,
title = "Model of the deviation of the medical needle during the movement in human tissue",
abstract = "In this paper, a mathematical model describing the deviation of the medical steel needle when it moves in the phantom of soft tissues (human tissues). This model will expect for use in the operational mode, where high precision positioning of the tip of the needle is necessary. The needle have asymmetric tip and because when the needle moves, it to deviate from rectilinear motion. Thus, by inserting and rotating the needle around its axis, it is possible to move the tip of the needle along a predetermined path. Consequently, having developed the model with its help, it will be possible to track the movement of the needle in the tissues or to predict optimal puncture points. The general formulation of the problems is considered and one of the sub-tasks is solved, as well as the available calculation of the needle tip deflection during the translational movement in a homogeneous material. The article shows the model used and the simulation results for different material densities and different needle tip angles. Comparison of the experimental data with the results obtained in the calculation using the proposed model is given.",
keywords = "Brachytherapy, Medical needle, Robotic system",
author = "Morozov, {Victor A.} and Nikitin, {Sergey A.} and Harlamov, {Vyacheslav V.} and Druzhinin, {Vasilii G.}",
year = "2018",
month = jan,
day = "1",
doi = "10.15593/RJBiomech/2018.4.03",
language = "English",
volume = "22",
pages = "401--413",
journal = "РОССИЙСКИЙ ЖУРНАЛ БИОМЕХАНИКИ",
issn = "1812-5123",
publisher = "Perm National Research Polytechnic University",
number = "4",

}

RIS

TY - JOUR

T1 - Model of the deviation of the medical needle during the movement in human tissue

AU - Morozov, Victor A.

AU - Nikitin, Sergey A.

AU - Harlamov, Vyacheslav V.

AU - Druzhinin, Vasilii G.

PY - 2018/1/1

Y1 - 2018/1/1

N2 - In this paper, a mathematical model describing the deviation of the medical steel needle when it moves in the phantom of soft tissues (human tissues). This model will expect for use in the operational mode, where high precision positioning of the tip of the needle is necessary. The needle have asymmetric tip and because when the needle moves, it to deviate from rectilinear motion. Thus, by inserting and rotating the needle around its axis, it is possible to move the tip of the needle along a predetermined path. Consequently, having developed the model with its help, it will be possible to track the movement of the needle in the tissues or to predict optimal puncture points. The general formulation of the problems is considered and one of the sub-tasks is solved, as well as the available calculation of the needle tip deflection during the translational movement in a homogeneous material. The article shows the model used and the simulation results for different material densities and different needle tip angles. Comparison of the experimental data with the results obtained in the calculation using the proposed model is given.

AB - In this paper, a mathematical model describing the deviation of the medical steel needle when it moves in the phantom of soft tissues (human tissues). This model will expect for use in the operational mode, where high precision positioning of the tip of the needle is necessary. The needle have asymmetric tip and because when the needle moves, it to deviate from rectilinear motion. Thus, by inserting and rotating the needle around its axis, it is possible to move the tip of the needle along a predetermined path. Consequently, having developed the model with its help, it will be possible to track the movement of the needle in the tissues or to predict optimal puncture points. The general formulation of the problems is considered and one of the sub-tasks is solved, as well as the available calculation of the needle tip deflection during the translational movement in a homogeneous material. The article shows the model used and the simulation results for different material densities and different needle tip angles. Comparison of the experimental data with the results obtained in the calculation using the proposed model is given.

KW - Brachytherapy

KW - Medical needle

KW - Robotic system

UR - http://www.scopus.com/inward/record.url?scp=85059617883&partnerID=8YFLogxK

U2 - 10.15593/RJBiomech/2018.4.03

DO - 10.15593/RJBiomech/2018.4.03

M3 - Article

AN - SCOPUS:85059617883

VL - 22

SP - 401

EP - 413

JO - РОССИЙСКИЙ ЖУРНАЛ БИОМЕХАНИКИ

JF - РОССИЙСКИЙ ЖУРНАЛ БИОМЕХАНИКИ

SN - 1812-5123

IS - 4

ER -