The use of π-flippers for correction of line integrals in Larmor-precession devices is described. The π-flippers consist of magnetised foils of the proper thickness, magnetised by the field of the precession device itself. Application of such π-flippers in a spin-echo small angle neutron scattering set-up with fields perpendicular and parallel to the beam direction is discussed. It appears that such flippers avoid corrections for the inhomogeneity of the field-line integrals, for the main part, in both options.
Язык оригинала | английский |
---|---|
Страницы (с-по) | S94-S96 |
Журнал | Applied Physics A: Materials Science and Processing |
Том | 74 |
Номер выпуска | SUPPL.I |
DOI | |
Состояние | Опубликовано - дек 2002 |