DOI

  • Pavel Buldakov
  • Natalia Resnina
  • Andrey Chumachenko
  • Matvey Stepanov
  • Pedro Del Nido
  • Olga Loboda
  • Nikolay Vasilyev

The work investigated functional and mechanical properties of the NiTi alloy specimens subjected to various thermal treatments at different temperatures. The stress-strain diagrams were studied at different temperatures. The results showed that a sample of the NiTi alloy wires of Ti - 50.84 at.% Ni alloys with a diameter of 0.5 mm that was annealed at 773 K for 1 hour demonstrated the best functional properties. This specimen showed good recovered strain 6 - 8%, low critical stress 250 - 450. Therefore, it was the best candidate for fabrication of an implantable device for repair of mitral valve prolapse.

Язык оригиналаанглийский
Страницы (с-по)4884-4888
Число страниц5
ЖурналMaterials Today: Proceedings
Том4
Номер выпуска3PB
DOI
СостояниеОпубликовано - 2017

    Предметные области Scopus

  • Материаловедение (все)

ID: 9332070