DOI

A method of tuning (scanning) X-ray beam wavelength based on modulation of the lattice parameter of X-ray optical crystal by an ultrasonic standing wave excited in it has been proposed and experimentally implemented. The double-crystal antiparallel scheme of X-ray diffraction, in which an ultrasonic wave is excited in the second crystal, is used in the experiment. The profile of characteristic line kα1 of an X-ray tube with a molybdenum anode is recorded using both the proposed tuning scheme and conventional mechanical rotation of crystal. The results obtained by both techniques are in good agreement.

Язык оригиналаанглийский
Страницы (с-по)170-172
Число страниц3
ЖурналCrystallography Reports
Том61
Номер выпуска2
DOI
СостояниеОпубликовано - 1 мар 2016
Опубликовано для внешнего пользованияДа

    Предметные области Scopus

  • Химия (все)
  • Материаловедение (все)
  • Физика конденсатов

ID: 88204480