Lithium-selective electrodes with solvent polymeric membranes based on tris [(N,N-dicyclohexylamide) neutral ionophore have been prepared and studied in conventional design (with an internal aqueous solution) and in solid-contact design with polyaniline as an ion-to-electron transducer. The dynamic range and the response slope of conventional and solid-contact electrodes are the same. One of the optional solid-contact designs studied provides a good long-term stability of readings. However, solid-contact electrodes exhibit lower selectivity to Li+ over H+ than conventional ISEs. The difference between the two kinds of electrodes is attributed to the pH sensitivity of polyaniline.
Язык оригинала | английский |
---|---|
Страницы (с-по) | 551-555 |
Число страниц | 5 |
Журнал | Electroanalysis |
Том | 14 |
Номер выпуска | 7-8 |
DOI | |
Состояние | Опубликовано - 2002 |