Optimum parameters were selected for determining the total content (TC) of Freon R-12 (CCl2F2) from ground spectrometric measurements of solar radiation at the St. Petersburg NDACC station on a Bruker FS125HR Fourier interferometer. On the basis of the obtained optimum set of parameters the total content of R-12 over the station was established over the period of 2009-2018 with a systematic error of 2.7% and a random error of 2.5% over the whole period of the investigations. The preliminary trend estimate was -0.46 +/- 0.06% per year.

Translated title of the contributionМетод обращения спектров прозрачности для оценки содержания CCl2F2 в атмосфере
Original languageEnglish
Pages (from-to)449-456
Number of pages8
JournalJournal of Applied Spectroscopy
Volume86
Issue number3
DOIs
StatePublished - Jul 2019

    Scopus subject areas

  • Condensed Matter Physics
  • Spectroscopy

    Research areas

  • freon content in the atmosphere, freon R-12, ground-based transparency method, ACE-FTS, CFC-11 CCL3F, FTIR MEASUREMENTS, EMISSIONS, VALIDATION, RETRIEVAL, HF

ID: 45498742