Fractures of the forearm. / Solomin, Leonid Nikolaevich; Kulesh, Pavel Nikolaevich.
The Basic Principles of External Skeletal Fixation Using the Ilizarov and Other Devices. Springer Nature, 2012. p. 339-373.Research output: Chapter in Book/Report/Conference proceeding › Chapter › Research › peer-review
}
TY - CHAP
T1 - Fractures of the forearm
AU - Solomin, Leonid Nikolaevich
AU - Kulesh, Pavel Nikolaevich
PY - 2012/1/1
Y1 - 2012/1/1
N2 - In Ilizarov external fixation of the forearm (Table 1.2), wires with a diameter of 1.5 mm are used. In combined external fixation, the basic wires have a diameter of 1.8–2 mm and the half-pins a diameter of 4 mm. The osteosynthesis set must also include 2-mm console wires with a stop set at various lengths corresponding to the part to be inserted into the bone (5, 10, 15, and 20 mm).
AB - In Ilizarov external fixation of the forearm (Table 1.2), wires with a diameter of 1.5 mm are used. In combined external fixation, the basic wires have a diameter of 1.8–2 mm and the half-pins a diameter of 4 mm. The osteosynthesis set must also include 2-mm console wires with a stop set at various lengths corresponding to the part to be inserted into the bone (5, 10, 15, and 20 mm).
UR - http://www.scopus.com/inward/record.url?scp=84956988895&partnerID=8YFLogxK
U2 - 10.1007/978-88-470-2619-3_11
DO - 10.1007/978-88-470-2619-3_11
M3 - Chapter
AN - SCOPUS:84956988895
SN - 9788847026186
SP - 339
EP - 373
BT - The Basic Principles of External Skeletal Fixation Using the Ilizarov and Other Devices
PB - Springer Nature
ER -
ID: 49736720